How to Tighten a
Loose Focusing Mechanism on the Substage Condenser
of an ML Series Microscope
|
|
Remove the condenser
from its mount by loosening the front thumbscrew
(red arrow) while holding the condenser
so it doesn't drop.
|
|
|
Using a
small Phillips screwdriver, remove the
screws holding the black covers on the
side of the substage.
|
|
|
Using a
2mm allen wrench, loosen the top and
bottom screws on the front of the
substage.
|
|
|
Using a
1.5mm allen wrench, tighten the top and
bottom screws on the side of the
substage until the condenser rack is no
longer loose and doesn't fall down.
|
|
|
Using a
2mm allen wrench, tighten the top and
bottom screws on the front of the
substage. Replace the black cover to
finish.
|
|